General information:
- This system is multi-function SCFE equipment, the working mode is intermittent or continuous
way to extract solvent from solid or liquid material, the co-solvent can be add to the supercritical
fluid;
- The system has two or more extractors and separators;
- Designed pressure of extractor is 40Mpa, max working pressure is 35Mpa, max working
temperature is 75;
- The design of pressure vessels according to the related national standard and rule demand and with
over pressure protection system which means enough safety;
- The material of extractors, separators, pipe and fitting, etc is all made of stainless steel 304.
- Safety protection device:
1. High-pressure pump outlet distribution contact pressure gauge, set the working pressure, overpressure automatic pump shutdown protection.
2. Extraction kettle, separation kettle, etc. according to the highest working pressure, respectively with safety valve, overpressure automatic pressure relief protection.